AP138
Part No.: AP138
Power Supply Type: Medical Power Supply
Special Feature:

1.For patient monitors, single output.

2.High efficiency,low temperature-rise, high quality, low ripple and noise.

3.Output current-limiting/over-voltage/voltage/short-circuit protection,etc.

 

Electrical characteristics
Input Output
Rated voltage 100~240vac Voltage 15Vdc
Voltage range 90~264Vac Current 4.3A
Frequency 47~63Hz Regulation ±5%
Efficiency ≥80% Ripple and noise ≤150mVp-p
Leakage current ≤3.5mA
Safety characteristics
Safety standard
Safety UL60601,EN60601
EMC CE EN550011,CLASS B
RE EN55011,CLASSB
Withstand voltage
Input-output Withstand voltage 4000Vac/5mA/1min
Input-FG Withstand voltage 2000Vac/5mA/1min
Input-Signal Withstand voltage
Output-FG Withstand voltage 500Vac/5mA/1min
Output-Signal Withstand voltage
Operating Environment
Operating Temperature 0℃~45℃
Relative Humidity 10%~95%
Storage temp. -40℃~+80
Cooling Natural convection
Structure drawing